● Appling principle:
ICH Harmonised Tripartite Guideline Q1A_R2 “Stabiltiy Testing of New Drug Substances and Products”
● Long term work ability:
It can continuously and non-stoply work for 2 years and it can automatically defrost.
Model | SHH-150SD | SHH-250SD | SHH-500SD | SHH-800SD | SHH-1000SD | ||
Test area size (mm) (D×W×H) | 500×500×600 | 600×550×750 | 720×720×950 | 850×750×1200 | 900×850×1300 | ||
Capacity | 150L | 250L | 500L | 765L | 1000L | ||
External size (mm) (D×W×H) | 880×680×1240 | 1000×680×1550 | 1140×860×1740 | 1270×880×1860 | 1300×1020×1980 | ||
Temperature range | 0 ~ 80℃ | ||||||
Temperature fluctuation | ±0.5℃ | ||||||
Humidity range | 20 ~ 98%R.H. | ||||||
Humidity fluctuation | ±2%R.H | ||||||
Operation mode | Fix | ||||||
Power supply | 220V,50Hz, 1P or 110V,50Hz, 1P | ||||||
Power | 1.5kw | 2kw | 2.5kw | 3.5kw | |||
Appling Guideline | ICH Harmonized Tripartite Guideline Q1A(R2) Stability Testing of New Drug Substances and Products |
External material | Printed cold-rolled steel plate |
Test area material | SUS304 stainless steel |
Isolation material | Superfine glass wool |
Window | Automatically defrosting window on the door |
Controller | LED Touch Screen or Digit display Intellectual temperature & humidity controller |
Operation mode | Fix and Constant |
Humidity sensor | Humidity transmitter |
Temp. sensor | PT100 temperature sensor |
Printer | A printer on the controlling panel. It continuously prints test data, including actual temperature, target temperature, actual humidity, target humidity and test time. |
USB (optional) | USB through which test data is continuously downloaded into U disk |
RS232 (optional) | RS232 interface through which the user may control, monitor and view test data on computer |
Compressor | "SECOP" hermetic compressor imported from Denmark |
Cooling mode | Air cooling |
Condenser | Domestically-made condenser |
Evaporator | Domestically-made evaporator |
Other cooling parts | Electric valve imported from Denmark |
Humidifier | Built-in stream boiler |
Water system | Built-in plastic water tank and Automatic water pump |
Heater | Nickel-cadmium heater |
Safety protection | Overheat, leakage, overpressure, overload, shout circuit protector |
Mobile alarm (optional) | Abnormal test data mobile phone alarm |
Other parts | Diameter ф50mm or ф100mm test hole |
Special Service | 3Q (IQ\PQ\OQ) verification documents |